img

Positive Pressure Isolator

This systems maintain an ISO Class 5 class in the Main Chamber during operation. This unit is used for carrying out sterile operations using a Laminar Flow system. The Isolator can be designed to handle operations like Sterile manufacturing, compounding, filtration, filling etc. It can also be used to house the complete injectable line.
This Isolator can be used to envelop Sterility Manufacturing Machines. The Sterility manufacturing machines will be in Client's scope, but will be integrated & programmed by IsoSaan.

Type : Rigid Positive Pressure
Air Flow Pattern : Laminar at 0.45 m/s
Operating Pressure : + 60 to + 100 Pascals
Operating Class : ISO Class 5
OEL Handling Range : OEL 4 to OEL 5
RTP : OEL 5
In-Feed Pass Box : With Sliding Trays
SART Port : With Disposable Gamma Treated Inserts
Pressurised Transfer : Using Nitrogen or Vacuum
RTP : OEL 5
START Port : OEL 5
  • Dedicated Inlet & Exhaust Blowers
  • Differential Pressure Transmitters across all filters & chambers
  • Temperature & RH Sensors
  • H2O2 Concentration Sensors
  • Optional 15” FLP IPC with SCADA
  • In-house System’s Validation & PLC validations
  • Automatic Leak Test
  • On Board Vapour used Hydrogen Peroxide Systems (VHP)
  • VHP Compatible Sensors
  • High & Low H2O2 Sensors
  • Catalyst Converter for diffusion of H2O2 before environmental discharge
  • Air Velocity Sensors
  • Viable & Non-Viable Particle Counters
  • Once through & re-circulatory systems
  • On Board Glove Integrity Test
  • Complete operational controls & feed backs of system pressure, filter health, door position, valve position & all parameters of Sterility equipment’s
  • Sterility Testing
  • Sterility Filling
  • Sterility Filtration
  • Sterility Powder Suspension
  • Sterility Injectable
  • Sterility Vile filling
  • Sterile Pouch Filling

Positive Pressure Isolator Manufacturers, Suppliers, Navi Mumbai, India